HPLC |
| Uptisphere, the different C18 Chemistries |
| Selectivity Guide |
|
Bonding |
C18-NEC |
C18-HDO |
C18-ODB |
C18-HSC |
C18-TF |
|
Code USP |
L1 |
L1 |
L1 |
L1 |
L1 |
|
- |
|||||
|
Pore size |
120A |
n.a. |
n.a. |
||
|
Surface area |
320 m2/g |
n.a. |
n.a. |
||
|
Bonding technology |
Monofunctional |
Trifunctional |
|||
|
End-capping |
no |
n.a. |
yes |
yes |
yes |
|
Carbon Content |
16% |
17% |
18% |
20% |
14% |
|
- |
|||||
|
3µm |
- |
x |
x |
x |
- |
|
5µm |
x |
x |
x |
x |
x |
|
7µm |
x |
- |
x |
- |
- |
|
10µm |
x |
x |
x |
- |
- |
|
15µm |
x |
x |
x |
- |
- |
|
20µm |
x |
- |
x |
- |
- |
|
Other
polar interactions can therefore be observed : silanophilic interactions
which are developed between the zeaxanthine hydroxyl and The higher this selectivity value is, the lower the interactions are ; the stationnary phase,therefore, carries fewer polar groups capable of interacting with the zeaxanthine, which has a low retention. This selectivity varies from 0.3 for phases with an inserted polar group, to almost 20 for the most protected phases. From 0.3 to 1 : the access to polar sites is high, from 1 to 5 it is average and from 5 to 10 it is low, above this it is very low. VariSous end-capping treatments (uniques or combined) lead most of the time to an increased selectivity, therefore showing the covering of residual silanols. Other non-end-capped phases with a high bonding density show a high selectivity it's means a reduced access to the polar sites of the stationnary phase. |
|
The hydrophobicity values cannot be shown on the same graph, they are given separately in the form of a histogramme.
|
|
|
The combination of the 3 probes allows you to define a set of groups, columns belonging to the same group will exhibit similar chromatographic results, even if the chemistry in support manufacture differs. |
|
|
Courtesy of E. Lesellier et A. Tchapla |
Selectivity 13 cis/all-trans b-carotene Courtesy of E. Lesellier et A. Tchapla |
|
For more information on the results of this test, the name of the different columns tested and their positions on the graph, please contact : LETIAM : letiam.colonnes@iut-orsay.fr INTERCHIM: interchrom@interchim.com Your local distributor network |
|