Advantages of using smaller particle size


  • Improved Efficiency
  • Lower Collection Volume
  • Higher Loading Capacity

Phthalates purification on 25 g Silica column
(125 mg Dimethyl Phthalate + 125 mg Diethyl Phthalate)
in 1 ml injection
Cyclohexane - Ethyl Acetate (85% - 15 %)
Flow rate : 30 ml/min
UV : 254nm

TOP

Flow rate impact on peak resolution


Gradient elution :
A) Hexane B) Ethyl acetate
0 - 100% B in 7 minutes
12 g column at 17, 30, 45, 60 ml/min
Methoxy acetophenone, Methyl paraben

TOP

Achieve greater Productivity @ Higher Flow rate with Smaller Particles


Phthalates purification on 25 g silica column
(125 mg Dimethyl Phthalate + 125 mg Diethyl Phthalate)
in 1 ml injection
Cyclohexane - Ethyl Acetate Gradient
100 % CH : 1 min
80 % CH : 7 min
80 % CH : 13 min

TOP

Selection Guide for Silica Columns


Analysis conditions :
Eluent : cyclohexane / ethyl acetate 90/10 - Flow rate : 40 ml/min

  Std type -
Irregular Silica 40/63 µm 40 g column
HP type -
High Performance Silica 50 µm 40 g column
HC type -
High Capacity Silica 50 µm 40 g column
Load 0
5 ml of solution A
Load 1
2 ml of solution C
Load 2
5 ml of solution C
Load 3
10 ml of solution C
Solution A Weight Volume Solvent
tri tertio butyl benzene 3 g 100 ml Cyclohexane / ethyl acetate 90 /10
Diethyl phthalate 3 g 100 ml
Dimethyl phthalate 3 g 100 ml
Solution C Weight Volume Solvent
tri tertio butyl benzene 2 g 100 ml Cyclohexane / ethyl acetate 90 /10
Diethyl phthalate 40 g 100 ml
Dimethyl phthalate 40 g 100 ml

TOP

Higher purity, in 7 minutes or 24 minutes - How much do you have ?


50 µm puriflash Silica Std, 25 g
P/N : IR-50SI/25G
FlowRate : 20 ml/min
Pressure : 0 bar
Solvents : A : Hexane - B : Ethyl acetate
Gradient CV : 0.0 - 9.6 CV, O - 20% B
Sample : 2 ml, 500 mg/ml, 4% Load
50 µm puriflash Silica HP, 25 g
P/N : IR-50SIHP/25G
FlowRate : 20 ml/min
Pressure : 0 bar
Solvents : A : Hexane - B : Ethyl acetate
Gradient CV : 0.0 - 4,8 CV, O - 20% B ; 4.8 - 5.6 CV, 20% B
Gradient time : 0:00 - 12:00,0 - 20% B ; 12:00 - 15:00, 20% B
Sample : 2 ml, 500 mg/ml, 4% Load
30 µm puriflash Silica HP, 25 g
P/N : PF-30SIHP/25g
FlowRate : 30 ml/min
Pressure : 1 bar
Solvents : A : Hexane - B : Ethyl acetate
Gradient CV : 0.0 - 4.8 CV, O - 20% B;4.8 - 5.6 CV, 20% B
Gradient time : 0:00 - 8:00,0 - 20% B ; 8:00 - 10:00, 20% B
Sample : 2 ml, 500 mg/ml, 4% Load
15 µm puriflash Silica HP, 25 g
P/N : PF-15SIHP/25G
FlowRate : 40 ml/min
Pressure : 2 bars
Solvents : A : Hexane - B : Ethyl acetate
Gradient CV : 0.0 - 4.8 CV, O - 20% B ;4.8 - 5.6 CV, 20% B
Gradient time : 0:00 - 6:00,0 - 20% B ; 6:00 - 7:00, 20% B
Sample : 2 ml, 500 mg/ml, 4% Load

TOP

Higher Capacity - 20 % sample load, 5 g of sample on a 25 g column !


50 µm puriflash Silica HC, 25 g
P/N : IR-50SIHC/25G
FlowRate : 20 ml/min
Pressure : 0 bar
Solvents : A : Hexane - B : Ethyl acetate
Gradient CV : 0.0 - 4.8 CV, O - 20% B ; 4.8 - 6.4 CV, 20% B
Gradient time : 0:00 - 12:00,0 - 20% B ; 12:00 - 16:00, 20% B
Sample : 5 ml, 500 mg/ml, 2.5 g, 10% Load
50 µm puriflash Silica HP, 25 g
P/N : IR-50SIHC/25G
FlowRate : 20 ml/min
Pressure : 0 bar
Solvents : A : Hexane - B : Ethyl acetate
Gradient CV : 0.0 - 4,8 CV, O - 20% B ; 4.8 - 6.6 CV, 20% B
Gradient time : 0:00 - 12:00,0 - 20% B ; 12:00 - 16:00, 20% B
Sample : 7 ml, 500 mg/ml, 3.5 g, 14% Load
50 µm puriflash Silica HC, 25 g
P/N : PF-50SIHC/25g
FlowRate : 20 ml/min
Pressure : 0 bar
Solvents : A : Hexane - B : Ethyl acetate
Gradient CV : 0.0 - 4.8 CV, O - 20% B;4.8 - 6.4 CV, 20% B
Gradient time : 0:00 - 12:00,0 - 20% B ; 12:00 - 16:00, 20% B
Sample : 10 ml, 500 mg/ml, 5.0 g, 20% Load

TOP


MONTLUCON - 211 bis avenue J.F.Kennedy
BP 1140 - 03103 Montluçon cedex
Tél. (33) (0) 4 70 03 88 55 - Fax (33) (0) 4 70 03 82 60
www.interchim.com - interchim@interchim.com