Flash Chromatography

 

IntelliFlash™ 280 System

Minimize Your Flash Purification Set-up time and Maximize Your Results

The IntelliFlashTM 280 Flash Purification System is a complete Flash purification platform designed for users that demand minimum set-up time, powerful functionality, and maximum results. Its modular hardware architecture, built-in WindowsÒ based control, scalable functions, and compact design will optimize flash purification efficiency and maximize your research effectiveness.

IntelliFlash System Features

IntelliFlashTM 280 system has an integrated modular hardware architecture with built-in computer control, binary gradient pump, sample injection module, separation modules, UV detector, and fraction collector

  • IntelliFlowTM twin piston pump with check valve free design for accurate positive displacement with automatic pressure monitoring & alarm for over/underpressure conditions

  • Hi volume fraction collection with FracTracTM to correlate fractions with control screen chromatogram

  • Customizable UV detector (254nm standard)

  • UniMountTM flexible flash module mounting system accepts any compression module/cartridge combination

  • EZ-LoadTM cartridge/module design with no insert/remove force

  • IntelliPeakTM software control and data collection with Wizard-based method programming software for quick and consistent method development

  • Modify method/run on the fly to improve purification performance

  • Extend method/run automatically to collect final peak purification

  • Only collect peak purifications

  • View digital chromatograph display and color coded corresponding fraction collection

  • Split fraction rack design allows continuous method run

  • Data storage on compact flash or PC cards, optional Ethernet communications, and email

280 System part numbers

ECL - 280 base model (comes with fraction collector and pump

UV Detector and Recorder
1- Include
0- Do not include

Gradient Formation Pump
1- Include
0- Do not include

System Components
A- F12 Compression Module
B- F25 Compression Module
C- F40 Compression Module
D- F65 Compression Module
E- 10 mL Sample Injection Module
F- 25 mL Sample Injection Module
G- 35 mL Sample Injection Module

 

Specifications

Description

Purification ch.

1

Solvant Delivery

Low pulsationvalveless dualpiston pump

Sequential purification ch.

1 standard/6 rnaximum

Gradient (# of solvents)

1 standard/2 maximum

Flow rate (mL/min.)

2-125 mL/min. (0.25% accuracy), Software flow adjustment

Maximum Pressure (PSI)

100 PSI (software low/high cut-off alarm.& noise filter)

Cartridge sizes

4-330g (F12, F25, F40, F75, F150)

Sample load size

5mg-33g (pre-packed inline sample loading)

Liquid sample injection

5-35mL, glass syringe

UV Detection

254nm std., 17 wavelength options

Fraction Tube sizes

13,16,18, 25mm Racks,bottles, scintilation

Max. Absorbance (OD)

400 (4.0000AU)

Control System

Integrated/embedded system, MS Windows CE OS

Processor

Intel StrongARM SA-1110 206MHz RISC

Video

VGA 6.5" TFT touch screen, 640 x 480, 64K colors, solvent resistent

Input

Compact keyboard with built in mouse control & Touch Screen nav.

Storage

32MB DRAM internal, 32MB Compact Flash (CF) card(64, 128, 256MB optional) USB,Optional PCMCIA

Ports

CF Card on front panel, Ethernet,USB on rear

Programming

Setup ‘Wizard’ guides user method development

File system

Standard Windows extended filenames & folders

Sample ID

Each method can be run repeatedly w/ different sample numbers

File Export

"Results Package" includes CSV data, Chromatogram JPEG & RTF (Word) Method configuration Excel (97/2000/XP) template reads & manipulates results pkg.

Connectivity

Ethernet comm. (TCP/lPnetwork & email "Results Pkg")

Auto Config.

Pre-run summary - ID resources for run (solvent racks, cartridges)

Chromatogram display

ln real-time w/historical pan/zoom/scroll by time & AU

Power

100 -260V 50/60 Hz (autoranging)

Weight

70 lbs (32 Kg)

Dimensions

25"H x 14"W x 16"D (65 x 36 x 40 cm)

Security

Removable flash (CF) media

 

211 Bis Avenue J.F. Kennedy - BP 1140 - 03103 Montluçon cedex - France

Tel. (33)(0)4 70 03 88 55  - Fax (33)(0)4 70 03 82 60 - Hot line: (33)(0)4 70 03 73 09

interchrom@interchim.com - www.interchim.com